SYSCAL R2
Deep Resistivity / IP System
Syscal R2 resistivity meters
Features
-
Powerful Electrical Resistivity
System
-
Entry Level IP System
-
Up to 1900 W Power
-
Output Current = 2.5A
-
Output Voltage = 1600V peak to peak
-
20 Bit A/D Converter
-
Compatible with Automated Electrode
Switching System
General
A unique configuration places the R2
resistivity meter in a class by itself. On one hand it
is a powerful DC resistivity system that may be powered
by external 12V or motor generator. On the other it is a
very portable, easy to use IP system. In either mode it
is a very accurate system featuring the Syscal
resistivity meter two channel design that permits
simultaneous measurement of both voltage and current.
The R2 system consists of three
components; the combined transmitter/receiver unit; a
DC/DC or AC/DC converter; and a power source, either
12 V battery, or motor generator.
The transmitter/receiver features a noise
monitoring system for pre-induction control consisting
of a DC digital voltmeter function. A line check/ground
resistance measurement allows the operator to check that
all electrodes are properly connected and grounded. Low
pass analog filters reduce the effect of higher
frequency natural and cultural noise.
The R2 resistivity meters will monitor
the input noise signal between the two receiving
electrodes so that noisy conditions may be noted. At the
start of each measurement the SP value is displayed.
This value is also stored with each reading. SP
compensation is automatic, with linear drift correction,
using digital filtering. The SP value is recalculated at
every third stack, which provides excellent correction
of the SP drift. electrical resistivity
The R2 resistivity meters have automatic
gain ranging for both voltage and current. Digital
stacking is used for noise reduction, with a maximum of
250 stacks. During stacking the operator can monitor
either the instantaneous reading, or the cumulative
average value.
Resolution after stacking is 1µV. The
standard deviation of each reading is also displayed and
stored. The R2 calculates apparent resistivity based on
the type of array and geometrical parameters that have
been entered. Up to 1,000 readings may be stored. Data
transfer is by serial link.
In the frequency mode the R2 has an
output waveform of ON+, ON-. In the time domain mode,
for resistivity and IP measurements, the waveform
becomes ON+, OFF, ON-, OFF. Pulse duration may be set
from 0.25-10 seconds.
The IP effect may be measured in four
windows whose width, and delay time, may be programmed.
The average value of the four chargeability windows is
calculated and displayed during measurement. At the end
of the measurement the chargeability for each window is
displayed, and may be stored. Chargeability values
normalized with respect to a standard decay curve can be
displayed for the detection of EM coupling or noise
effects.
The second component of a Syscal R2
electrical resistivity system is the power converter.
This may be either a DC/DC converter, or a motor
generator powered AC/DC converter. At the lower power
range is a 125W DC/DC converter, powered by external
12V. With a marginal increase in weight the next step up
is a 250W DC/DC converter that is still powered by a 12 V
battery. The final step is a motor generator powered
AC/DC converter with maximum 1,200W output. The third
component making up the system is the power source
itself, either 12V, or 220V; either battery or motor
generator supplied.
The Syscal R2 resistivity meter is truly
a unique system able to handle the deepest resistivity
applications, to depths of 2,000 meters or more. It is
also a very portable IP system suitable for many shallow
mineral exploration programs. The R2 may be utilized
with the Intelligent Node automated switching system.
Specifications
Power Sources
The Syscal R2 resistivity meters must be
powered by an external DC source for ground energization.
Available sources:
A 125W DC/DC converter supplied by a 12V
battery. Ranges: 50V - 2.50A max, 100V -1.25A max, 200V
-0.62A max, and 400V - 0.31 A max. Dimensions: 17 x 21 x
28cm. Weight: 4kg.
A
250W DC/DC converter supplied by a 12V battery.
Ranges: 100V - 2.50A max, 200V 1.25A max, 400V -0.62A
max, and 800V - 0.31 A max. Dimensions: 31 x 21 x 2 1
cm. Weight: 5kg.
A 1200W AC/DC converter supplied by a
220V (or 110V) motor generator. Ranges: 50V - 1.5A max,
100V - 1.5A max, 200V - 1.5A max, 400V 1.5A max, and
800V -1.5A max. Dimensions: 43 x 29 x 32cm. Weight:
25kg.
Induced Polarization Measurements
Beyond apparent resistivity, the Syscal
R2 resistivity meters also measures electrical
resistivity using the Time Domain Induced Polarization
(IP) parameter. The IP method is used in mineral
exploration for the detection of massive and scattered
mineralization.
A pulse-current waveform is transmitted
(ON+, OFF, ON-, OFF) and the IP delay curve is analyzed
with four changeability windows. The pulse duration (ON
time), the delay time and the width of each individual
window are programmable. The average value (M) of the
four partial chargeabilities (Mi) is computed and
displayed during measurement, while the four Mi values
are available after measurement. resistivity
meters
Chargeability values normalized with
respect to a standard decay curve can be displayed for
the detection of EM coupling or noise effects. With this
versatile chargeability measurement, the Syscal R2
functions as both a resistivity meter and a low-weight,
low-cost IP system for shallow mineral exploration.
Measured Parameters
-
Measurement and display of current,
voltage, standard deviation, and SP
-
Computation and display of apparent
resistivity for main electrode arrays: Schlumberger,
Wenner, Gradient, and dipole-dipole.
-
Measurement and display of IP
chargeability with four windows. Delay time and
window widths individually programmable.
-
Display of noise level before
measurement
-
Measurement and display of ground
resistance (electrode check)
Memory Operations
-
Storage of data in the internal
memory: Up to 1022 readings. Each reading includes
current, voltage, SP, resistivity, chargeability,
spacing, station number, etc.
-
Reading of data stored in the memory
on the LCD display
-
Transfer of data to a computer
(ASCII, binary files) through RS-232 serial link
-
Direct transfer of data to a printer
-
Remote control of the unit by a
computer through the serial link.
Transmitter
-
Maximum output voltage: 800V (1600V
peak to peak)
-
Maximum Output Current: 2A (2.5A in
option)
-
Maximum Output Power: 1600W (1900W in
option)
-
External DC source for ground
energization (DC/DC or AC/DC converters)
-
Output Current Resolution: 10µA.
Accuracy. Standard 0.3%, max 1% from -20°C to + 70°C
-
Output Current Waveform: Frequency
domain ON+, ON- for resistivity. Time domain ON+,
OFF, ON-, OFF for resistivity and chargeability.
Pulse duration (ON time) programmable from 0.25 to
10 s, with preset values of 0.5 s, 1 s and 2 s.
-
Thermal circuit breaker for
overheating protection
Receiver:
-
Input Impedance: 10 Mohms
-
Input over voltage protection
-
Input Voltage Range: - 5V to + 5V
-
Automatic SP bucking (-5V to + 5V)
with linear drift correction up to 1 mV/s
-
50Hz and 60Hz power line rejection
-
Ground Resistance Measurement: 0.1 to
1,000k ohms
-
Voltage Measurement: Resolution: 1µV
after stacking
-
A/D Converter: 20 Bit
-
Accuracy: Standard 0.3%, max 1% from
-20°C to +70°C
-
Chargeability Measurement -
Resolution: 0. 1mV/V Accuracy: 1% of displayed value
for a voltage greater than l0mv
-
Continuous digital stacking up to 250
stacks
Miscellaneous
-
LCD display of 2 lines of 20
characters
-
Weather-resistant case
-
Dimensions: 31 x 21 x 21cm
-
Weight: 6kg, including dry cells
-
Power supply for electronic circuits:
Six 1.5V D size dry cells. Over 300 hours operation
at 20°C
-
External DC source for ground
energization (DC/DC or AC/DC converters)
-
Operating Temperature Range: -20°C to
+ 70°C (Optionally -40°C to + 70°C)
-
Storage Temperature: -40°C to +80°C
Standard Components
Syscal R2 resistivity meter console,
battery charger, data transfer cable, software, transit
case and instruction manual.
|
 |
 |
 |
 |
 |
|
 |