SYSCAL R2 Deep Resistivity
/ IP System

Features
-
Powerful Electrical Resistivity
System
-
Entry Level IP System
-
Up to 1900 W Power
-
Output Current = 2.5A
-
Output Voltage = 1600V peak to
peak
-
20 Bit A/D Converter
-
Compatible with Automated
Electrode Switching System
General
A unique configuration places the
R2 resistivity meter in a class by itself. On one hand
it is a powerful DC resistivity system that may be
powered by external 12V or motor generator. On the other
it is a very portable, easy to use IP system. In either
mode it is a very accurate system featuring the Syscal
resistivity meter two channel design that permits
simultaneous measurement of both voltage and current.
The R2 system consists of three
components; the combined transmitter/receiver unit; a
DC/DC or AC/DC converter; and a power source, either 12
V battery, or motor generator.
The transmitter/receiver features a
noise monitoring system for pre-induction control
consisting of a DC digital voltmeter function. A line
check/ground resistance measurement allows the operator
to check that all electrodes are properly connected and
grounded. Low pass analog filters reduce the effect of
higher frequency natural and cultural noise.
The R2 resistivity meters will
monitor the input noise signal between the two receiving
electrodes so that noisy conditions may be noted. At the
start of each measurement the SP value is displayed.
This value is also stored with each reading. SP
compensation is automatic, with linear drift correction,
using digital filtering. The SP value is recalculated at
every third stack, which provides excellent correction
of the SP drift.
The R2 resistivity meters have
automatic gain ranging for both voltage and current.
Digital stacking is used for noise reduction, with a
maximum of 250 stacks. During stacking the operator can
monitor either the instantaneous reading, or the
cumulative average value.
Resolution after stacking is 1µV.
The standard deviation of each reading is also displayed
and stored. The R2 calculates apparent resistivity based
on the type of array and geometrical parameters that
have been entered. Up to 1,000 readings may be stored.
Data transfer is by serial link.
In the frequency mode the R2 has an
output waveform of ON+, ON-. In the time domain mode,
for resistivity and IP measurements, the waveform
becomes ON+, OFF, ON-, OFF. Pulse duration may be set
from 0.25-10 seconds.
The IP effect may be measured in
four windows whose width, and delay time, may be
programmed. The average value of the four chargeability
windows is calculated and displayed during measurement.
At the end of the measurement the chargeability for each
window is displayed, and may be stored. Chargeability
values normalized with respect to a standard decay curve
can be displayed for the detection of EM coupling or
noise effects.
The second component of a Syscal R2
electrical resistivity system is the power converter.
This may be either a DC/DC converter, or a motor
generator powered AC/DC converter. At the lower power
range is a 125W DC/DC converter, powered by external
12V. With a marginal increase in weight the next step up
is a 250W DC/DC converter that is still powered by a 12
V battery. The final step is a motor generator powered
AC/DC converter with maximum 1,200W output. The third
component making up the system is the power source
itself, either 12V, or 220V; either battery or motor
generator supplied.
The Syscal R2 resistivity meter is
truly a unique system able to handle the deepest
resistivity applications, to depths of 2,000 meters or
more. It is also a very portable IP system suitable for
many shallow mineral exploration programs. The R2 may be
utilized with the Intelligent Node automated switching
system.
Specifications
Power Sources
-
The Syscal R2 resistivity
meters must be powered by an external DC source for
ground energization. Available sources:
-
A 125W DC/DC converter supplied
by a 12V battery. Ranges: 50V - 2.50A max, 100V
-1.25A max, 200V -0.62A max, and 400V - 0.31 A max.
Dimensions: 17 x 21 x 28cm. Weight: 4kg.
-
A
250W DC/DC
converter supplied by a 12V battery. Ranges:
100V - 2.50A max, 200V 1.25A max, 400V -0.62A max,
and 800V - 0.31 A max. Dimensions: 31 x 21 x 2 1 cm.
Weight: 5kg.
-
A 1200W AC/DC converter
supplied by a 220V (or 110V) motor generator.
Ranges: 50V - 1.5A max, 100V - 1.5A max, 200V - 1.5A
max, 400V 1.5A max, and 800V -1.5A max. Dimensions:
43 x 29 x 32cm. Weight: 25kg.
Induced Polarization
Measurements
Beyond apparent resistivity, the
Syscal R2 resistivity meters also measures electrical
resistivity using the Time Domain Induced Polarization
(IP) parameter. The IP method is used in mineral
exploration for the detection of massive and scattered
mineralization.
A pulse-current waveform is
transmitted (ON+, OFF, ON-, OFF) and the IP delay curve
is analyzed with four changeability windows. The pulse
duration (ON time), the delay time and the width of each
individual window are programmable. The average value
(M) of the four partial chargeabilities (Mi) is computed
and displayed during measurement, while the four Mi
values are available after measurement.
Chargeability values normalized
with respect to a standard decay curve can be displayed
for the detection of EM coupling or noise effects. With
this versatile chargeability measurement, the Syscal R2
functions as both a resistivity meter and a low-weight,
low-cost IP system for shallow mineral exploration.
Measured Parameters
-
Measurement and display of
current, voltage, standard deviation, and SP
-
Computation and display of
apparent resistivity for main electrode arrays:
Schlumberger, Wenner, Gradient, and dipole-dipole.
-
Measurement and display of IP
chargeability with four windows. Delay time and
window widths individually programmable.
-
Display of noise level before
measurement
-
Measurement and display of
ground resistance (electrode check)
Memory Operations
-
Storage of data in the internal
memory: Up to 1022 readings. Each reading includes
current, voltage, SP, resistivity, chargeability,
spacing, station number, etc.
-
Reading of data stored in the
memory on the LCD display
-
Transfer of data to a computer
(ASCII, binary files) through RS-232 serial link
-
Direct transfer of data to a
printer
-
Remote control of the unit by a
computer through the serial link.
Transmitter
-
Maximum output voltage: 800V
(1600V peak to peak)
-
Maximum Output Current: 2A
(2.5A in option)
-
Maximum Output Power: 1600W
(1900W in option)
-
External DC source for ground
energization (DC/DC or AC/DC converters)
-
Output Current Resolution:
10µA. Accuracy. Standard 0.3%, max 1% from -20°C to
+ 70°C
-
Output Current Waveform:
Frequency domain ON+, ON- for resistivity. Time
domain ON+, OFF, ON-, OFF for resistivity and
chargeability. Pulse duration (ON time) programmable
from 0.25 to 10 s, with preset values of 0.5 s, 1 s
and 2 s.
-
Thermal circuit breaker for
overheating protection
Receiver
-
Input Impedance: 10 MOhms
-
Input over voltage protection
-
Input Voltage Range: - 5V to +
5V
-
Automatic SP bucking (-5V to +
5V) with linear drift correction up to 1 mV/s
-
50Hz and 60Hz power line
rejection
-
Ground Resistance Measurement:
0.1 to 1,000k ohms
-
Voltage Measurement:
Resolution: 1µV after stacking
-
A/D Converter: 20 Bit
-
Accuracy: Standard 0.3%, max 1%
from -20°C to +70°C
-
Chargeability Measurement -
Resolution: 0. 1mV/V Accuracy: 1% of displayed value
for a voltage greater than l0mv
-
Continuous digital stacking up
to 250 stacks
Miscellaneous
-
LCD display of 2 lines of 20
characters
-
Weather-resistant case
-
Dimensions: 31 x 21 x 21cm
-
Weight: 6kg, including dry
cells
-
Power supply for electronic
circuits: Six 1.5V D size dry cells. Over 300 hours
operation at 20°C
-
External DC source for ground
energization (DC/DC or AC/DC converters)
-
Operating Temperature Range:
-20°C to + 70°C (Optionally -40°C to + 70°C)
-
Storage Temperature: -40°C to
+80°C
Standard Components
Syscal R2 resistivity meter
console, battery charger, data transfer cable, software,
transit case and instruction manual.
|