HFP-2293 Heat-Pulse Flow Meter

Theory of Operation
The Heat Pulse flow meter operation is as follows. The
tool is lowered into the borehole via a cable attached
to a winch. When the tool is in position to take a flow
measurement the trigger assembly button is pressed. This
sends a pulse down the center conductor which when
detected by circuitry in the probe, fires the heat grid
and signals the surface monitoring equipment and
software to begin a flow measurement cycle. The grid
heats ambient fluids and if there is up or down flow in
the well, this heated fluid mass is detected at
thermistor sensors (2 cm from grid) allowing the time
(and flow rate) through a known x-sectional area to be
recorded by an amplifier. The output of this amplifier
is then converted to a frequency. This frequency is then
driven up the cable line and monitored by the surface
equipment. When the tool is pulsed by the surface
system, the tool immediately begins to charge the
capacitors that produce the voltage for the heat grid in
preparation for the next measurement cycle. A complete
flow measurement is made when the time is accurately
measured from when the heat grid is fired to when a peak
temperature change, carried by the flow, is detected by
either the upper or lower sensor.
Software
MSHeat is the Windows acquisition software for the
HFP-2293. This software is compatible with MGX II
systems running MSLog. Individual heat-flow waveforms
can be saved, and text files with depth and flow rate
can be imported into WellCAD for a histogram-type
presentation.
Specifications
0.113. litters/minute to 3.785
litters/minute (0.03 gallons/minute to 1.0
gallons/minute)
0.046. meters/minute. to 3.962 meters/minute (0.15
feet/minute to 13 feet/minute)
-
Resolution:. 5%
-
Accuracy:. 5% (Mid-Range) to
15% (Extremes)
-
Dimensions:. Length:. 122 cm
(48 inches)
-
Diameter:. 4.1 cm (1.63 inches)
-
Weight:. 5.5 kg (12 lbs)

|