AB-5 Portable Radon Monitor

Features
-
Programmable sampling.
-
Data storage.
-
Continuous mode.
-
Grab samples.
-
Printer option.
-
RS-232-C option.
-
Rechargeable battery.
-
Impact resistant housing.
-
Real time clock.
-
pCi/L output.
General
The Model AB-5 Portable Radiation
Monitor is a versatile lightweight instrument for fast
accurate measurement of environmental radiation levels.
The unit is completely self-contained and has a built-in
photomultiplier tube, air pump and rechargeable battery
to minimize setup time. A wide variety of probes and
detectors are available as accessories. Users can
purchase a basic instrument and add accessories as
required.
Accessories are available for
measuring alpha or beta emitting particulates on a 47mm
diameter filter A standard ABA adapter, and a continuous
passive radon and environmental level radon gas detector
enable the operator to measure alpha emitting gases by
various methods. Wipe testing and surface contamination
accessories are simple attachments. Gamma-beta and low
energy gamma and x-ray probes plug into the rear of the
instrument.
The AB-5 radiation monitor has many
applications. For environmental measurements, the
instrument offers a truly portable means of collecting
field data on a programmable and long term basis. The
instrumentation is not limited to detection of a single
class of radiation and is therefore an excellent general
purpose lab instrument. Accessories are available for
health protechon surveys or environmental monitoring.
This includes wipe testing and surface contamination
surveys.
Specifications
Applications
Continuous monitoring, grab
sampling, radon gas, radon decay products, environmental
studies, area scanning, field surveys, surface
contamination, airborne particulates, and x-ray leakage.
AB-5 Portable Radon Monitor
Operating Modes: Continuous, radon,
quasi-continuous and ratemeter
Background Rate: 0.4 counts per
minute
Maximum Count Rate: 106 counts per
second
Maximum Number of Runs: 81
Maximum Number of Intervals: 675
Interval Length: 1 to 99 seconds,
minutes, or hours
Variable Pump: 0 to 3.7 I/min
Variable Photomultiplier Tube: 600
to 1 100vdc
Battery: Internal rechargeable
Maximum Operating Time*: Internal
pump off- 75 hours (nominal), internal pump on (at 3.7
Vmin. continuous)- 8 hours (nominal)
Operating Temperature: -10°C to
50°C
Dimensions (without handle): 21.6 x
9.6 x 24. lcm
Printer (optional): PPT-1 portable
printer Parallel, dot matrix
* For extended operation an
external battery kit is available
CPRD Continuous Passive Radon
Detector
Sensitivity: 1.50 cpm/pCi/1 nominal (0.040 cpm/Bq/m3)
LLD** (1 hour count): 0.3 pCi/1(11
Bq/m3)
Background: 0.5cpm nominal
Scintillator: ZnS(Ag)
Active Volume: 272ml
Dimensions: 55 x 15.2cm Weight:
188g
Model AEP-47,and BEP - 7 Detectors
|
|
AEP-47 |
BEP-47 |
|
Rad. Detected: |
alpha |
beta |
|
Energy Range: |
4.5 to 9.0 MeV |
0.1 to 3.0 MeV |
|
Filter Size: |
47 mm |
47 mm |
|
Scintillator: |
ZnS(Ag) |
plastic BC-00 |
|
Nom. Efficiency: |
16% |
16% |
|
Max Bkgrnd: |
0.5cpm |
15cpm |
|
Dimensions: |
62 x 72 mm |
62 x 72 mm |
|
Weight: |
200g |
200g |
Model PMT-SA, SB, and SAB Contamination Probes
|
|
PMT-SA |
PMT-SB |
PMT-SAB |
|
Rad. Detected: |
alpha |
beta |
alpha or beta |
|
Sensitive Area: |
45.5cm² |
45.5cm² |
45.5cm² |
|
Alpha Energy: |
4.5 to 9.0 |
|
4.5 to 9.0 |
|
Ranges: |
MeV |
|
MeV |
|
Beta Energy Ranges: |
|
0.1 to 3.0 MeV |
0.1 to 3.0 MeV |
|
Scintillator: |
ZnS(Ag) |
plastic
BC-400
1mm thick |
plastic
BC-400
1mm thick |
|
Alpha Efficiency: |
20% |
|
20% |
|
Beta Efficiency: |
|
14% |
13% |
|
Background: |
0.5cpm |
30cpm |
30cpm |
|
Weight: |
630 g |
630g |
630g |
Note: SA, SB, and SAB contamination probes require PMT
PMT-TEL Trace Environmental. Level Radon Gas Detector
Sensitivity: 23 cpm/pCi/1 (0.622 cpm/Bq/m³) nominal
LLD** (24 hour count): 0.005 pCi/1 (0.19 Bq/m³) (radon)
Active Volume: 17.5 liters nominal
Background: 0.5cpm nominal
Scintillator: ZnS(Ag)
Operating Temperature: 0°C to 40°C
Battery Type: Four "C" alkaline cells and one 9V alkaline battery. Battery Iife 1000 hours
Dimensions: 480 x 370mm, Weight: 8.0kg
PMT-EL Environmental Level Radon Gas Detector
Sensitivity: 7.0 cpm/pCi/1 (0.189 cpm/Bq/m³) nominal
LLD** (24 hour count): 0.02 pCi/l (0.74 Bq/m³) (radon)
Active Volume: 5 liters
Background: 0.5cpm nominal
Scintillator: ZnS(Ag)
Operating Temperature: 0°C to 40°C
Battery Type: Four "AA" alkaline cells
Battery Life: 300 hours
Dimensions: 220 x 330mm. Weight: 2.15kg
Model WT-A, B, and AB Wipe Test Assemblies
|
|
WT-A |
WT-B |
WT-AB |
|
Rad. Detected: |
alpha |
beta |
alpha and beta |
|
Alpha Energy: |
4.5 to 9.0 MeV |
|
4.5 to 9.0 MeV |
|
Beta Energy: |
|
0.1 to 3.0 MeV |
0.1 to 3.0 MeV |
|
Scintillator: |
ZnS(Ag) |
plastic
BC-400 |
ZnS(Ag)
plastic
BC-400 |
|
Alpha Efficiency: |
24% |
|
24% |
|
Beta Efficiency: |
|
16% |
16% |
|
Background: |
0.5cpm |
15cpm |
15cpm |
|
Dimension: |
76 x 46mm |
76 x 46mm |
76 x 46mm |
|
Weight: |
150g |
150g |
150g |
Model 110A, and 300A Lucas Cells
|
|
110A |
300A |
|
Volume: |
151 ml |
270 ml |
|
Scintillator: |
ZnS(Ag) |
ZnS(Ag) |
|
Active Area: |
18,800 mm² |
27,700 mm² |
|
Counting Efficiency: |
0.745±.02 cpm/dpm |
0.745±.02 cpm/dpm |
|
Sensitivity: |
0.75 cpm/pCi/l
(0.02cpm/Bq/m³)
nominal |
1.37 cpm/pCi/1
(0.037 cpm/Bq/m³)
nominal |
|
Dimensions: |
58 x 127.5mm |
58 x 180mm |
|
LLD**
(1 hour count): |
0.6pCi/l
(22 Bq/m³) |
0.3pCi/l
(11 Bq/m³) |
** Lower limit of detection
Standard Components
AB-5 portable radon monitor with
ABA adaptor, built-in sampling pump, battery and battery
charger, and instruction manuals.
|